Add to Cart
ASTM Standard UV Accelerated Aging Test Chamber With Programmable Controller
HD-E802 UV Aging Chamber - UVA.pdf
Introduction:
UV Accelerated Aging Test Chamber can simulate the sun, rain and dew to damage the material, UV machine using UV lamp simulate the effect of sunlight, cold rain and dew. Being tested materials are placed in a certain environment which the light and moisture are alternating.
The materials are tested for several days or weeks, you can know the damage of material at outside for
several months or several years, Damage types include: bleaching, discoloration, loss of light, powder,
crack, turbidity, bubble-based, embrittlement, strength decline and oxidation.
Standard Features:
Item | Specification |
Chamber dimension (W*D*H) | 1340*530*1400mm |
Chamber material | SUS#304 stainless steel |
Temperature range | RT to 70℃ |
Temperature fluctuation | ≤±1.0℃ |
Temperature uniformity | 3℃ |
Humidity range: | ≥75%RH |
Controller | Programmable controller, LCD touch screen |
Control mode | Balance temperature humidity control (BTHC) |
Test cycle setting | Exposure, condensation and water spray test cycle is programmable |
Water spray cycle | Spray 18mins in every 120mins / spray 12mins in every 60mins |
Water usage | 8L/day |
Distance from sample to lamp | 55±2mm |
Centre distance between lamps | 65~70mm |
Irradiance range | 0.45~0.90W/m2 |
UV lamps | Imported Atlas UV-A: 315-400nm (8pcs, 1600h lifetime) |
Lamp power | 40W/Piece |
Specimen Size | 75×290mm (24pcs) / 75x150mm (48pcs), maximum thickness 5mm |
Testing time | 0~999H, adjustable |
Protection system | Overload short circuit protection Over temperature protection Water lacking protection Earth leakage protection Auto shut off protection |
Applications
Control System:
1.The blackboard temperature meter controls the heating and the temperature is more stable.
2.Radiometer probes (optional) are fixed and do not have to be loaded or unloaded every time.
3.A dedicated ultraviolet irradiation meter that uses high-precision display and measurement of radiation.
4.Lighting and condensation can be independently controlled and can be alternately cycled.
Separate control time for lighting and condensation and alternating cycle control can be set arbitrarily within one thousand hours
Test Criteria:
ISO 4892-3
ASTM G154